Display
|
LCD, 3 1/2digit; 1999 Digits
|
Display ranges
|
|
Earth bond resistance
|
0.05...19.99 Ω
|
Test current / voltage
|
> 200 mA AC/approx. 19 V AC
|
Insulation resistance
|
0.1...19.99/199.9/250 MΩ
|
Test voltage / current
|
500 V DC/> 1 mA
|
Substitute leakage current
|
0.25...19.99 mA
|
Substitute housing leakage curent
|
0.25...19.99 mA
|
Substitute patient leakage current
|
0.25...19.99 mA
|
Test voltage/current
|
approx. 50 V AC/<3.5mA
|
Touch leakage current
|
0.1...1.99 mA
|
Trip Voltage
|
50 V DC...500 V DC
31 V AC...312 V AC
|
General
|
|
Data memory
|
for approx. 600 measurement values equal to approx. 150 UUT
|
Interface
|
IR-RS-232
|
Limit display
|
symbols in LCD
|
Safety complying with
|
DIN VDE 0404/
IEC 61010/EN 61010/
DIN VDE 0411/
IEC 61557/EN 61557/
DIN VDE 0413
|
Measurement category
|
CAT II/300 V
|
Pollution degree
|
2
|
Protection class
|
II
|
Protection degree
|
IP 40
|
Power supply
|
230 V +10%/-15%, 50 Hz
|
Dimensions
|
105 x 295 x 70 mm
|
Weight
|
pprox. 1.3 kg
|